Z-1506-2022 Class II Ongoing

Recalled by Guangdong Haiou Medical Apparatus Co., Ltd — Puning

FDA device recall Z-1506-2022 was initiated by Guangdong Haiou Medical Apparatus Co., Ltd on June 10, 2022 and is designated Class II. Reason for recall: Complaints received in which needle and needle holder have detached from the syringe after injection as well as needle not retracting into the syringe as per intended use, resulting in an increased ri… The recall status is ongoing. Affected quantity: 4,050,000 units.

Recall Details

Product Type
Devices
Report Date
August 17, 2022
Initiation Date
June 10, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4,050,000 units

Product Description

HAIOU Needle retractable safety syringe with needle-(1ml syringe with 25G x 1-inch needle} 200 pieces/box. Intended Use: Disposable sterile needle retractable safety syringe. Catalog Number: NRSS1-25G25

Reason for Recall

Complaints received in which needle and needle holder have detached from the syringe after injection as well as needle not retracting into the syringe as per intended use, resulting in an increased risk of needle stick injury to the end user

Distribution Pattern

US Nationwide distribution in the state of CA.

Code Information

Lot Number: 20IB0 Exp. Date: 2025/09/04