Z-1514-2019 Class II Terminated

Recalled by Maquet Cardiovascular, LLC — Wayne, NJ

FDA device recall Z-1514-2019 was initiated by Maquet Cardiovascular, LLC on Apr 18, 2019 and is designated Class II. Reason for recall: Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path… The recall status is terminated (terminated Jul 28, 2025).

Recall details

Product Type
Devices
Report Date
Jun 5, 2019
Initiation Date
Apr 18, 2019
Termination Date
Jul 28, 2025
Voluntary/Mandated
Voluntary: Firm initiated

Product description

Maquet Getinge-BEQ-TOP 15705 Material: 701051122

Reason for recall

Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured

Distribution pattern

Nationwide

Code information

Batch Numbers: 3000044566 3000046394 3000049202 3000054550 3000058043 3000060239 3000066276 3000069368 3000071833 3000074339 3000080701 3000081598