Z-1516-2014 Class II Terminated
Recalled by Acumed LLC
— Hillsboro, OR
FDA device recall Z-1516-2014 was initiated by Acumed LLC on Mar 26, 2014 and is designated Class II. Reason for recall: Certain lots of Acumed Tension Band Pins are being recalled because the implantable product was manufactured with a grade of stainless steel that is not within specifications. The recall status is terminated (terminated Feb 4, 2015). Affected quantity: 2544 unit in the US and 3584 units outside the US.
Recall details
- Product Type
- Devices
- Report Date
- May 7, 2014
- Initiation Date
- Mar 26, 2014
- Termination Date
- Feb 4, 2015
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2544 unit in the US and 3584 units outside the US
Product description
The Acumed Tension Band Pin is a fixation pin which is .0625" in diameter and is available in lengths between 35mm and 90mm. It is to be used in conjunction with commonly used orthopedic wire. 50.0 mm Tension Band Pin; Part Number 30-0097 70.0 mm Tension Band Pin; Part Number 30-0098 90.0 mm Tension Band Pin; Part Number 30-0099 50.0 mm Tension Band Kit; Part Number TB-1550K-S 70.0 mm Tension Band Kit; Part Number TB-1570K-S 90.0 mm Tension Band Kit; Part Number TB-1590K-S Product Usage: Acumed Tension Band Pin is intended to be used with orthopedic wire to address malleolar, patella, and olecranon fractures in tension band wiring procedures. This device is not intended for usage in the spine.
Reason for recall
Certain lots of Acumed Tension Band Pins are being recalled because the implantable product was manufactured with a grade of stainless steel that is not within specifications.
Distribution pattern
US Nationwide Distribution - in the states of CA, CO, FL, GA, LA, MI, MN, MO, OH, OR, PA, PR, TX, UT, VA, WA and WY. Product was distributed outside the US to: Australia, China, Figi, France, Great Britain, Italy, Japan, Korea, Maylasia, Sweden, South Africa, Spain.
Code information
50.0 mm Tension Band Pin: Lot 243639 70.0 mm Tension Band Pin: Lots 241127, 242707, 243325, 243948. 90.0 mm Tension Band Pin: Lots 241128. 242708, 243327, 243949. 50.0 mm Tension Band Kit: Lots 216769, 217002, 217941, 219145, 219402, 222732, 223007, 224312, 229889, 232575, 235490, 237772, 244800, 246204, 247056. 70.0 mm Tension Band Kit: Lots 210290, 211349, 214574, 215405, 217004, 218273, 219230, 220978, 221503, 223008, 223477, 224066, 225554, 227193, 229394, 230287, 230952, 231971, 233742, 235502, 237773, 240512, 241185, 243617, 243945, 244801, 246777, 246779, 246938, 247108, 247109, 247493. 90.0 mm Tension Band Kit: Lots 215053, 217005, 218600, 218700, 221465, 222731, 223009, 223756, 225555, 228951, 231573, 236969, 237774, 243620, 244802, 246939, 246990, 247381.
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1516-2014
- openFDA file date
- Nov 19, 2026
- Loaded by RegDataLab
- Nov 20, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.