Z-1520-2013 Class II Terminated

Recalled by DePuy Orthopaedics, Inc. — Warsaw, IN

FDA device recall Z-1520-2013 was initiated by DePuy Orthopaedics, Inc. on May 3, 2013 and is designated Class II. Reason for recall: The Glenosphere Orientation Guide Instrument Used with the Delta XTEND Reverse Shoulder is being recalled because an arrow that is etched on the instrument was incorrectly placed on the opposite side… The recall status is terminated (terminated Aug 28, 2013). Affected quantity: 20 units.

Recall Details

Product Type
Devices
Report Date
Jun 19, 2013
Initiation Date
May 3, 2013
Termination Date
Aug 28, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
20 units

Product Description

DePuy Glenosphere orientation guide Product Usage: The glenosphere orientation guide is an instrument used in reverse shoulder arthroplasty. The orientation guide is used to properly align the glenosphere.

Reason for Recall

The Glenosphere Orientation Guide Instrument Used with the Delta XTEND Reverse Shoulder is being recalled because an arrow that is etched on the instrument was incorrectly placed on the opposite side of the instrument. Using the improperly etched arrow as a guide could potentially lead to incorrect orientation of the glenosphere implant.

Distribution Pattern

Worldwide Distribution - USA Nationwide and countries of: Australia, Canada, Czech Republic, England, Germany, Switzerland and Turkey.

Code Information

Part Number 230795000/Lot Number 5120443

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1520-2013
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.