FDA device recall Z-1522-2014 was initiated by Integra LifeSciences Corp. on Mar 10, 2014 and is designated Class II. Reason for recall: Some 23 kHz Torque Wrenches were incorrectly assembled and contained a 36 kHz Hex Wrench Insert instead of the 23 kHz Hex Wrench Insert. The recall status is terminated (terminated Feb 13, 2015). Affected quantity: 3,984 devices packaged as 6 devices / box (664 boxes).
Recall details
- Product Type
- Devices
- Report Date
- May 7, 2014
- Initiation Date
- Mar 10, 2014
- Termination Date
- Feb 13, 2015
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 3,984 devices packaged as 6 devices / box (664 boxes)
Product description
Integra CUSA Excel Disposable Wrench. Catalog Number C5601. The CUSA Excel+ Sterile 23 kHz Torque Wrench is an accessory used with the CUSA Excel+ System ultrasonic surgical instrument. The torque wrench is used to properly secure the CUSA tip to the CUSA handpiece in the sterile field.
Reason for recall
Some 23 kHz Torque Wrenches were incorrectly assembled and contained a 36 kHz Hex Wrench Insert instead of the 23 kHz Hex Wrench Insert.
Distribution pattern
Worldwide Distribution - USA (nationwide) including the countries of Australia, Canada, and Switzerland.
Code information
Product/Catalog Number C5601; Lot Numbers 1132424, 1132883, 1132884, 1133453, 1133454, 1133455, 1134067, 1134675, 1134873 and 1135014