Z-1522-2014 Class II Terminated

Recalled by Integra LifeSciences Corp. — Plainsboro, NJ

FDA device recall Z-1522-2014 was initiated by Integra LifeSciences Corp. on Mar 10, 2014 and is designated Class II. Reason for recall: Some 23 kHz Torque Wrenches were incorrectly assembled and contained a 36 kHz Hex Wrench Insert instead of the 23 kHz Hex Wrench Insert. The recall status is terminated (terminated Feb 13, 2015). Affected quantity: 3,984 devices packaged as 6 devices / box (664 boxes).

Recall details

Product Type
Devices
Report Date
May 7, 2014
Initiation Date
Mar 10, 2014
Termination Date
Feb 13, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3,984 devices packaged as 6 devices / box (664 boxes)

Product description

Integra CUSA Excel Disposable Wrench. Catalog Number C5601. The CUSA Excel+ Sterile 23 kHz Torque Wrench is an accessory used with the CUSA Excel+ System ultrasonic surgical instrument. The torque wrench is used to properly secure the CUSA tip to the CUSA handpiece in the sterile field.

Reason for recall

Some 23 kHz Torque Wrenches were incorrectly assembled and contained a 36 kHz Hex Wrench Insert instead of the 23 kHz Hex Wrench Insert.

Distribution pattern

Worldwide Distribution - USA (nationwide) including the countries of Australia, Canada, and Switzerland.

Code information

Product/Catalog Number C5601; Lot Numbers 1132424, 1132883, 1132884, 1133453, 1133454, 1133455, 1134067, 1134675, 1134873 and 1135014

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1522-2014
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1522-2014. Report a data issue: hello@regdatalab.com.