Z-1523-2014 Class I Terminated
Recall Details
- Product Type
- Devices
- Report Date
- May 28, 2014
- Initiation Date
- March 21, 2014
- Termination Date
- March 20, 2018
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 11,881 units (US 5183 and OUS 6698)
Product Description
Datascope System 98/98xt, CS 100/CS100i and CS 300 Intra-Aortic Balloon Pump. The Intra-Aortic Balloon Pump (IABP) is an electromechanical system used to inflate and deflate intra-aortic balloons in cardiovascular procedures.
Reason for Recall
Potential mechanical failure of the fan assembly associated with the power supply. A fan assembly failure could result in the power supply overheating and cause the IABP to shut down without warning.
Distribution Pattern
Worldwide Distribution - USA (nationwide) and approximately 104 countries.
Code Information
IABP Model Numbers System 98 0998-00-0446-xx 0998-UC-0446-xx System 98XT 0998-00-0479-xx 0998-UC-0479-xx CS100i 0998-UC-0446Hxx 0998-UC-0479Hxx CS100 0998-00-3013-xx 0998-UC-3013-xx CS300 0998-00-3023-xx 0998-UC-3023-xx