Z-1523-2022 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- August 24, 2022
- Initiation Date
- July 19, 2022
- Termination Date
- September 23, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 45 devices
Product Description
Power electronic box, Part #AS-0066, subassembly for the MyoCycle MC-2.
Reason for Recall
Screws mounting an internal power supply may come loose during shipping resulting in increased risk of high voltage electrical shock.
Distribution Pattern
Distribution was made to AZ, CA, CT, FL, GA, ID, IL, KY, MA, MI, MN, MO, MN, NC, NJ, NM, NY, OH, OR, TX, UT, VA, WA, and WI. There was no foreign/military/government distribution.
Code Information
Serial numbers 000800 thru 000819, 000821 thru 000822, 000824 thru 000827, 000829 thru 000831, 000835, 000836, 000840 thru 000842, 000850 thru 000852, and 000857 thru 000859; UDI 000861553000417