Z-1530-2020 Class II Ongoing

Recalled by Karl Storz Endoscopy — El Segundo, CA

FDA device recall Z-1530-2020 was initiated by Karl Storz Endoscopy on Aug 27, 2019 and is designated Class II. Reason for recall: During a complaint investigation it was determined that an error was observed within the instruction manual included with some scopes. Specifically, The Quick Connector Kits (QCK) and tray… The recall status is ongoing. Affected quantity: 73 scopes.

Recall details

Product Type
Devices
Report Date
Apr 1, 2020
Initiation Date
Aug 27, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
73 scopes

Product description

Karl Storz NEURO-FIBERSCOPE, Model # 11282BN1, Flexible Pediatric Neuroscope

Reason for recall

During a complaint investigation it was determined that an error was observed within the instruction manual included with some scopes. Specifically, The Quick Connector Kits (QCK) and tray information, specified for the STERIS System 1E (SS1E) sterilization method within the instruction manual is incorrect. If the user uses SS1E to sterilize the impacted scopes and referred to the manual concerning the QCK information, the sterility of reprocessed scope cannot be guaranteed.

Distribution pattern

US: AL, AZ, CA, CO, CT, DE, FL, GA, IL, IN, MA, MD, MI, MN, MO, NC, NH, NJ, OH, OR, PA, SC, TN, TX, VA, WA, WI, WV, OUS: Canda

Code information

All Lots

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1530-2020
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Karl Storz Endoscopy? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Karl Storz Endoscopy: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Karl Storz Endoscopy →

Source: openFDA record for Z-1530-2020. Report a data issue: hello@regdatalab.com.