Z-1534-2017 Class II Terminated
FDA device recall Z-1534-2017 was initiated by Siemens Medical Solutions USA, Inc on Feb 22, 2017 and is designated Class II. Reason for recall: In Artis systems with A100G generators, a component can be affected by aging which could result in the potential failure of a module in the high-voltage generator. The recall status is terminated (terminated Oct 24, 2017). Affected quantity: 264 distributed worldwide (112 in the U.S.).
Recall details
- Product Type
- Devices
- Report Date
- Mar 29, 2017
- Initiation Date
- Feb 22, 2017
- Termination Date
- Oct 24, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 264 distributed worldwide (112 in the U.S.)
Product description
Reason for recall
Distribution pattern
Distributed throughout the United States.
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1534-2017
- openFDA file date
- Nov 27, 2026
- Loaded by RegDataLab
- Nov 28, 2026