Z-1534-2019 Class II Terminated

Recalled by Maquet Cardiovascular, LLC — Wayne, NJ

FDA device recall Z-1534-2019 was initiated by Maquet Cardiovascular, LLC on April 18, 2019 and is designated Class II. Reason for recall: Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path b… The recall status is terminated (terminated July 28, 2025). Affected quantity: N/A.

Recall Details

Product Type
Devices
Report Date
June 5, 2019
Initiation Date
April 18, 2019
Termination Date
July 28, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
N/A

Product Description

Maquet Getinge-BEQ-TOP 21503 ECC 1/4" AV LOOP Material:701054776

Reason for Recall

Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured

Distribution Pattern

Nationwide

Code Information

Batch Numbers: 3000043295 3000047248 3000047254 3000055638 3000062271 3000065551 3000069074 3000075386