Z-1536-2018 Class II Terminated
FDA device recall Z-1536-2018 was initiated by Smith & Nephew, Inc. on Dec 11, 2017 and is designated Class II. Reason for recall: Packaging design is insufficient to prevent potential sterile barrier breach (pin hole) in Tyvek film of pouch. A limited number of previously recalled devices were inadvertently redistributed. The recall status is terminated (terminated Sep 21, 2020). Affected quantity: 22.
Recall details
- Product Type
- Devices
- Report Date
- May 2, 2018
- Initiation Date
- Dec 11, 2017
- Termination Date
- Sep 21, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 22
Product description
Reason for recall
Distribution pattern
Distributed to accounts in 7 states: AZ, ME, NY, OH, PA, TN, and VT; and PR.
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1536-2018
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026