Z-1537-2019 Class II Terminated
FDA device recall Z-1537-2019 was initiated by Maquet Cardiovascular, LLC on Apr 18, 2019 and is designated Class II. Reason for recall: Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path… The recall status is terminated (terminated Jul 28, 2025).
Recall details
- Product Type
- Devices
- Report Date
- Jun 5, 2019
- Initiation Date
- Apr 18, 2019
- Termination Date
- Jul 28, 2025
- Voluntary/Mandated
- Voluntary: Firm initiated
Product description
Maquet Getinge-BEQ-T 8025 1/4" ECC Pack with RF32 Material:701055142
Reason for recall
Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured
Distribution pattern
Nationwide
Code information
Batch Numbers: 3000044427 3000054806 3000059172 3000069470
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1537-2019
- openFDA file date
- Nov 16, 2026
- Loaded by RegDataLab
- Nov 17, 2026