Z-1542-2013 Class II Terminated

Recalled by Edwards Lifesciences, LLC — Draper, UT

FDA device recall Z-1542-2013 was initiated by Edwards Lifesciences, LLC on May 14, 2013 and is designated Class II. Reason for recall: Edwards is recalling the QuickDraw Venous Cannula due to complaints of cannula separation during withdrawal of the device when the Percutaneous insertion method is used. The recall status is terminated (terminated Oct 14, 2014). Affected quantity: 22,616.

Recall details

Product Type
Devices
Report Date
Jun 19, 2013
Initiation Date
May 14, 2013
Termination Date
Oct 14, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
22,616

Product description

QuickDraw Venous Cannula with Insertion Kit. ThruPort Systmes. Models QD22 (22 Fr) and QD25 (25 Fr). The product is packaged sterile and non-pyrogenic in a sealed, peel-type pouch. Two pouches (cannula and accessories) go into one shelf carton. One shelf carton goes in one shipper carton. Use of the QuickDraw venous cannula is indicated for patients undergoing cardiopulmonary bypass. The QuickDraw venous cannula serves to drain non-oxygenated blood from the venae cavae or right atrium during cardiopulmonary bypass. The product is packaged sterile and nona¿ pyrogenic in a sealed, peel-type pouch.

Reason for recall

Edwards is recalling the QuickDraw Venous Cannula due to complaints of cannula separation during withdrawal of the device when the Percutaneous insertion method is used.

Distribution pattern

Worldwide distribution: US Nationwide and country of Canada.

Code information

Model QD22: Lot No. 59049758, 59054488, 59067502, 59074170, 59081626, 59083736, 59091652, 59094887, 59114755, 59117451, 59134508, 59149651, 59154310, 59178579, 59201243, 59212118, 59222657, 59233636, 59233637, 59246684, 59246701, 59255899, 59255900, 59269439, 59284780, 59287725, 59296462, 59299605, 59315339, 59320575, 59334652, 59356129, 59360072, 59365198, 59365199, 59374009, 59384309, 59390552, 59394166, 59394167, 59421250, 59448841 Model QD25: Lot No. 59046082,59046847,59052393,59052395,59056270,59059029,59060798,59062681,59069772,59073141,59077234,59079573,59085168,59085169,59097289,59101637,59106005,59110720,59122491,59126281,59131923,59140133,59142266,59143453,59145027,59158524,59158579,59163164,59184252,59191107,59201246,59207044,59209886,59214360,59214361,59216991,59222658,59222659,59233651,59246685,59246703,59255901,59268625,59284781,59287728,59296463,59299610,59313057,59320582,59320583,59334656,59344191,59346468,59352608,59356135,59360084,59374014,59374015,59414791,59425355,59437699,59437700,59448853

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1542-2013
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Edwards Lifesciences, LLC? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Edwards Lifesciences, LLC: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Edwards Lifesciences, LLC →

Source: openFDA record for Z-1542-2013. Report a data issue: hello@regdatalab.com.