Z-1543-2021 Class I Terminated

Recalled by STEP-HAR MEDICAL LLC — Fountain Valley, CA

Recall Details

Product Type
Devices
Report Date
May 19, 2021
Initiation Date
February 26, 2021
Termination Date
August 15, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2

Product Description

Alaris Pump Module, Model 8100, serviced/repaired by Step-Har Medical with an impacted bezel repair part

Reason for Recall

Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may result in free flow, over infusion, under infusion or interruption of infusion.

Distribution Pattern

IA

Code Information

Serial Numbers: 12747983 and 9919591