Z-1543-2024 Class II Ongoing

Recalled by Smiths Medical ASD Inc. — Minneapolis, MN

Recall Details

Product Type
Devices
Report Date
April 24, 2024
Initiation Date
December 19, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1842 guide barrel clamps

Product Description

Smiths Medical Medfusion Guide Barrel Clamp, Item number G6000716

Reason for Recall

Medfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead to slippage of the spring within the barrel clamp assembly. If this occurs, it could result in the inability of the pump to recognize a syringe or the pump may misidentify the size of the syringe loaded.

Distribution Pattern

Worldwide distribution.

Code Information

Lot Numbers: P0407365, P0486670, P0561740