Z-1552-2025 Class II Ongoing

Recalled by CareFusion 303, Inc. — San Diego, CA

Recall Details

Product Type
Devices
Report Date
April 23, 2025
Initiation Date
October 5, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2679

Product Description

MEDFLEX, REF: 1119-00; MEDBANK MAIN, REF: 1137-00; MEDFLEX 2.0, REF: 1139-00; MEDBANK TOWER MAIN, REF: 1145-00; MEDBANK MINI MAIN, REF: 1147-00 BD PYXIS MEDBANK Facility Software, REF: 139090-01; BD PYXS MEDBNK TWR MN 2HH1FH2HM2FM2FG1FF, REF: 169-100; BD PYXIS MEDBANK MINI 1HH-1FH- 1HM, REF: 169-101; BD PYXIS MEDBANK MINI 1HH-1HM- 1FM, REF: 169-102; BD PYXIS MEDBANK MINI 1HH-3HM, REF: 169-103; BD PYXIS MEDBANK MINI 2FH, REF: 169-104; BD PYXIS MEDBANK MINI 2FM, REF: 169-105; BD PYXIS MEDBANK MINI 2HH-1FH, REF: 169-106; BD PYXIS MEDBANK MINI 2HH-1FM , REF: 169-107; BD PYXIS MEDBANK MINI 2HH-2HM, REF: 169-108; BD PYXIS MEDBANK MINI 2HM-1FM, REF: 169-109; BD PYXIS MEDBANK MINI 3HH-1FG, REF: 169-110; BD PYXIS MEDBANK MINI 3HH-1HM, REF: 169-111; BD PYXIS MEDBANK MINI 4HH, REF: 169-112; BD PYXIS MEDBANK MINI 4HM, REF: 169-113; BD PYXIS MDBANK TWR MN 7HH-1HM-3FM RxNow, REF: 169-132; BD PYXIS MEDBANK MINI 1FH-1FM, REF: 169-137; BD PYXIS MEDBANK TWR MN 1FH- 10HM-1FM, REF: 169-138; BD PYXIS MEDBANK TWR MN 2HH- 4FH-2FM, REF: 169-139; OBS BD PYXIS MB TOWER MAIN, REF: 169-153; OBS BD PYXIS MB TWR RXNOW MAIN W/MED PRI, REF: 169-155; OBS BD PYXIS MB TWR RXNOW MAIN W/PAT PRI, REF: 169-156; OBS BD PYXIS MB MINI MAIN W/CUBIE DRWS, REF: 169-159; OBS BD PYXIS MB MINI MAIN W/MATRIX DRWS, REF: 169-160; OBS BD PYXIS MB MINI MAIN W/HH FALSE DRW, REF: 169-161; OBS BD PYXIS MB MINI MAIN W/FH FALSE DRW, REF: 169-162; BD PYXIS MEDBANK TWR MN 10HH- 1FH-1FM, REF: 169-23; BD PYXIS MEDBANK TWR MN 10HH- 2FH, REF: 169-24; BD PYXIS MEDBANK TWR MN 10HH- 2FM, REF: 169-25; BD PYXIS MEDBANK TWR MN 12HH- 1FF, REF: 169-26; BD PYXIS MEDBANK TWR MN 12HH- 1FH, REF: 169-27; BD PYXIS MEDBANK TWR MN 12HH- 1FM, REF: 169-28; BD PYXIS MEDBANK TWR MN 14HH, REF: 169-29; BD PYXIS MEDBANK TWR MN 1HH- 1HM-2FM-4FF, REF: 169-30; BD PYXIS MEDBANK TWR MN 2FH- 2HM-4FM, REF: 169-31; BD PYXIS MEDBANK TWR MN 2HH- 10HM-1FM, REF: 169-32; BD PYXIS MEDBANK TWR MN 2HH- 1FH-5FM, REF: 169-33; BD PYXIS MEDBANK TWR MN 2HH- 2FH-4FM, REF: 169-34; BD PYXIS MEDBANK TWR MN 2HH- 2FM-4FF, REF: 169-35; BD PYXIS MEDBANK TWR MN 2HH- 4HM-4FF, REF: 169-36; BD PYXIS MEDBANK TWR MN 2HH- 4HM-4FM, REF: 169-37; BD PYXIS MEDBANK TWR MN 2HH- 6FM, REF: 169-38; BD PYXIS MEDBANK TWR MN 2HH- 6HM-3FM, REF: 169-39; BD PYXIS MEDBANK TWR MN 2HH- 8HM-2FM, REF: 169-40; BD PYXIS MEDBANK TWR MN 2HM- 6FM, REF: 169-41; BD PYXIS MEDBANK TWR MN 3FH-4FM , REF: 169-42; BD PYXIS MEDBANK TWR MN 3HH- 7HM-2FM, REF: 169-43; BD PYXIS MEDBANK TWR MN 4HH- 10HM, REF: 169-44; BD PYXIS MEDBANK TWR MN 4HH- 1FH-2HM-3FM, REF: 169-45; BD PYXIS MEDBANK TWR MN 4HH- 1FH-4FM, REF: 169-46; BD PYXIS MEDBANK TWR MN 4HH- 1FH-4HM-2FM, REF: 169-47; BD PYXIS MEDBANK TWR MN 4HH- 1FH-6HM-1FM, REF: 169-48; BD PYXIS MEDBANK TWR MN 4HH- 2FH-2HM-2FM, REF: 169-49; BD PYXIS MEDBANK TWR MN 4HH- 2FH-3FM, REF: 169-50; BD PYXIS MEDBANK TWR MN 4HH- 2HM-4FM, REF: 169-51; BD PYXIS MEDBANK TWR MN 4HH- 3FH-2FM, REF: 169-52; BD PYXIS MEDBANK TWR MN 4HH- 4FH-1FM, REF: 169-53; BD PYXIS MEDBANK TWR MN 4HH- 4HM-3FM, REF: 169-54; BD PYXIS MEDBANK TWR MN 4HH-5FH, REF: 169-55; BD PYXIS MEDBANK TWR MN 4HH- 5FM, REF: 169-56; BD PYXIS MEDBANK TWR MN 4HH- 6HM-2FM, REF: 169-57; BD PYXIS MEDBANK TWR MN 4HH- 8HM-1FM, REF: 169-58; BD PYXIS MEDBANK TWR MN 4HM- 5FM, REF: 169-59; BD PYXIS MEDBANK TWR MN 5HH- 3HM-3FM, REF: 169-60; BD PYXIS MEDBANK TWR MN 5HH- 9HM, REF: 169-61; BD PYXIS MEDBANK TWR MN 6FH-1FM, REF: 169-62; BD PYXIS MEDBANK TWR MN 6HH- 1FH-2HM-2FM, REF: 169-63; BD PYXIS MEDBANK TWR MN 6HH- 1FH-3FM, REF: 169-64; BD PYXIS MEDBANK TWR MN 6HH- 2FH-2FM, REF: 169-65; BD PYXIS MEDBANK TWR MN 6HH- 2FH-2HM-1FM, REF: 169-66; BD PYXIS MEDBANK TWR MN 6HH-4FH, REF: 169-67; BD PYXIS MEDBANK TWR MN 6HH- 4FM, REF: 169-68; BD PYXIS MEDBANK TWR MN 6HH- 4HM-2FM, REF: 169-69; BD PYXIS MEDBANK TWR MN 6HH- 6HM-1FM, REF: 169-70; BD PYXIS MEDBANK TWR MN 6HH- 8HM, REF: 169-71; BD PYXIS MEDBANK TWR MN 7FH, REF: 169-72; BD PYXIS M

Reason for Recall

If "Clear All" selected medications is selected on patient profile, the automated dispensing cabinet may allow dispensing of medication without first requiring the RxVerify Validation Code be entered/without validation from a pharmacist, which could lead to unintended and incorrect medication removal.

Distribution Pattern

US Nationwide distribution in the states of KS, TX, CA, PA, NJ, FL, IN, AZ, IL, MA, NY, KY, UT, MD, WA, OH, AL.

Code Information

BD Pyxis MedBank Receiving Cabinet Software Software versions: 2.1.2.11, 2.2.1.8, 2.31.8, 2.3.2.1, 2.4.2.26 REF/UDI-DI: 1119-00/10885403512704; 1137-00; 1139-00/10885403512704; 1145-00/10885403517167; 1147-00/10885403517181; 139090-01; 169-100/10885403512568; 169-101/10885403512544; 169-102/10885403512544; 169-103/10885403512544; 169-104/10885403512544; 169-105/10885403512544; 169-106/10885403512544; 169-107/10885403512544; 169-108/10885403512544; 169-109/10885403512544; 169-110/10885403512544; 169-111/10885403512544; 169-112/10885403512544; 169-113/10885403512544; 169-132/10885403512568; 169-137/10885403512544; 169-138/10885403512568; 169-139/10885403512568; 169-153/10885403512568; 169-155/10885403512568; 169-156/10885403512568; 169-159/10885403512544; 169-160/10885403512544; 169-161/10885403512544; 169-162/10885403512544; 169-23/10885403512568; 169-24/10885403512568; 169-25/10885403512568; 169-26/10885403512568; 169-27/10885403512568; 169-28/10885403512568; 169-29/10885403512568; 169-30/10885403512568; 169-31/10885403512568; 169-32/10885403512568; 169-33/10885403512568; 169-34/10885403512568; 169-35/10885403512568; 169-36/10885403512568; 169-37/10885403512568; 169-38/10885403512568; 169-39/10885403512568; 169-40/10885403512568; 169-41/10885403512568; 169-42/10885403512568; 169-43/10885403512568; 169-44/10885403512568; 169-45/10885403512568; 169-46/10885403512568; 169-47/10885403512568; 169-48/10885403512568; 169-49/10885403512568; 169-50/10885403512568; 169-51/10885403512568; 169-52/10885403512568; 169-53/10885403512568; 169-54/10885403512568; 169-55/10885403512568; 169-56/10885403512568; 169-57/10885403512568; 169-58/10885403512568; 169-59/10885403512568; 169-60/10885403512568; 169-61/10885403512568; 169-62/10885403512568; 169-63/10885403512568; 169-64/10885403512568; 169-65/10885403512568; 169-66/10885403512568; 169-67/10885403512568; 169-68/10885403512568; 169-69/10885403512568; 169-70/10885403512568; 169-71/10885403512568; 169-72/10885403512568; 169-73/10885403512568; 169-74/10885403512568; 169-75/10885403512568; 169-76/10885403512568; 169-77/10885403512568; 169-78/10885403512568; 169-79/10885403512568; 169-80/10885403512568; 169-81/10885403512568; 139089-01