Z-1556-2019 Class II Terminated

Recalled by Maquet Cardiovascular, LLC — Wayne, NJ

FDA device recall Z-1556-2019 was initiated by Maquet Cardiovascular, LLC on April 18, 2019 and is designated Class II. Reason for recall: Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path b… The recall status is terminated (terminated July 28, 2025). Affected quantity: N/A.

Recall Details

Product Type
Devices
Report Date
June 5, 2019
Initiation Date
April 18, 2019
Termination Date
July 28, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
N/A

Product Description

Maquet Getinge-BEQ-T 5211 LARGE ECC SET Material:701064560

Reason for Recall

Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured

Distribution Pattern

Nationwide

Code Information

Batch Numbers: Batch 3000041128 3000045731 3000050577 3000052947 3000054617 3000054955 3000057631 3000071818