Z-1559-2025 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- April 23, 2025
- Initiation Date
- March 24, 2025
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 114,202 units
Product Description
Welch Allyn CONNEX Accessory Power Management Stand:
Reason for Recall
There were customer reports of devices which experienced battery-related fires. The investigation found that the devices in question were utilizing third-party, after-market batteries that are not approved by Baxter.
Distribution Pattern
Worldwide - US Nationwide distribution.
Code Information
All Serial Numbers. Product Code: 7000-APM, UDI-DI: 00732094210613; Product Code: 75-HCA-CTB, UDI-DI: 00732094240597; Product Code: 75-HCA-MTB, UDI-DI: 00732094240603;