Z-1559-2025 Class II Ongoing

Recalled by Baxter Healthcare Corporation — Deerfield, IL

Recall Details

Product Type
Devices
Report Date
April 23, 2025
Initiation Date
March 24, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
114,202 units

Product Description

Welch Allyn CONNEX Accessory Power Management Stand:

Reason for Recall

There were customer reports of devices which experienced battery-related fires. The investigation found that the devices in question were utilizing third-party, after-market batteries that are not approved by Baxter.

Distribution Pattern

Worldwide - US Nationwide distribution.

Code Information

All Serial Numbers. Product Code: 7000-APM, UDI-DI: 00732094210613; Product Code: 75-HCA-CTB, UDI-DI: 00732094240597; Product Code: 75-HCA-MTB, UDI-DI: 00732094240603;