Z-1563-2016 Class II Terminated

Recalled by 3M Company - Health Care Business — Saint Paul, MN

FDA device recall Z-1563-2016 was initiated by 3M Company - Health Care Business on Jan 22, 2016 and is designated Class II. Reason for recall: 3M discovered a valve with incorrectly labeled "input" and "output" orientation. This defect can result in the valve being installed backwards, creating a remote possibility of valve failure. A… The recall status is terminated (terminated Aug 25, 2016). Affected quantity: 74 valves.

Recall Details

Product Type
Devices
Report Date
May 4, 2016
Initiation Date
Jan 22, 2016
Termination Date
Aug 25, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
74 valves

Product Description

3M Steri-Vac Gas Sterilizer/Aerator, models 5XL and 8XL.

Reason for Recall

3M discovered a valve with incorrectly labeled "input" and "output" orientation. This defect can result in the valve being installed backwards, creating a remote possibility of valve failure. A failure may result in EO being vented outside the sealed chamber during a limited period in the EO injection phase of the cycle.

Distribution Pattern

AL, CO, CT, FL, GA, IN, NC, NJ, NY, OK, OR, PA, RI, VA, WI, and OUS to include: Argentina, Canada, Chile, China, Columbia, El Salvador, Germany, India, Indonesia, Japan, Korea, Mexico, Peru, Philippines, South Africa, Switzerland, Thailand, Trinidad, Turkey, United Arab Emirates, Vietnam.

Code Information

Model 5XL, Serial Numbers: 722612 - 722683, 820352 - 820360 Model 8XL, Serial Numbers: 351303 - 351399, 450144 - 450147

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1563-2016
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.