Z-1565-2014 Class II Terminated

Recalled by SpineFrontier, Inc. — Beverly, MA

FDA device recall Z-1565-2014 was initiated by SpineFrontier, Inc. on Dec 23, 2011 and is designated Class II. Reason for recall: It has been verified through field use that there are two possible breakages that may occur when using SI50001 Revision D Lock Shaft Inserter, possibly resulting in instrument fragments entering the… The recall status is terminated (terminated May 20, 2014). Affected quantity: 30 devices.

Recall details

Product Type
Devices
Report Date
May 14, 2014
Initiation Date
Dec 23, 2011
Termination Date
May 20, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
30 devices

Product description

SpineFrontier S-Lift Lock Shaft Inserter, an instrument used with S-LIFT Intervertebral Body Fusion Device in spinal fusion surgery. Product Usage: The S-Lift Lock Shaft SI50001 mates with SI50002, Implant Inserter, to aid in securing the implant to the whole implant inserter assembly, SI50000.

Reason for recall

It has been verified through field use that there are two possible breakages that may occur when using SI50001 Revision D Lock Shaft Inserter, possibly resulting in instrument fragments entering the sterile field: Impact Cap: The impact cap on the Lock Shaft Inserter can shear off under high-impact force during insertion and removal. Pin: The pin used to guide the Lock Shaft Inserter into the ma

Distribution pattern

US Nationwide - OR, FL, MA, TX, CO, OK, IL, and MO

Code information

S-Lift Lock Shaft Part number; SI50001 Revisions: A, B Lots: 42613, 2478901R, 42615, 7625701

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1565-2014
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1565-2014. Report a data issue: hello@regdatalab.com.