Z-1567-2020 Class I Ongoing

Recalled by LeMaitre Vascular, Inc. — Burlington, MA

FDA device recall Z-1567-2020 was initiated by LeMaitre Vascular, Inc. on Mar 23, 2020 and is designated Class I. Reason for recall: There has been an increased trend in reports of catheters failing to deflate during use, which may ultimately lead to tip separation. The recall status is ongoing. Affected quantity: 43315.

Recall details

Product Type
Devices
Report Date
Apr 8, 2020
Initiation Date
Mar 23, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
43315

Product description

LeMaitre 5F Plus Over the Wire Embolectomy Catheter, Model Nos. 1651-84, e1651-84, 1651-88, e1651-88. The LeMaitre Over the Wire Embolectomy Catheter is indicated for use in the removal of emboli and thrombi during embolectomy and/or thrombectomy. It can also be used for catheter placement over a guidewire, vessel occlusion, fluid infusion and/or aspiration.

Reason for recall

There has been an increased trend in reports of catheters failing to deflate during use, which may ultimately lead to tip separation.

Distribution pattern

Worldwide distribution: US (Nationwide distribution) and foreign countries of: Australia, Austria, Belgium, Bulgaria, Colombia, Costa Rica, Czech Republic, Dominican Republic, Egypt, El Salvador, Estonia, Finland, France, Germany, Georgia, Hungary, Ireland, Israel, Italy, Japan, Korea, Netherlands, New Zealand, Norway, Poland, Portugal, Qatar, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, United Kingdom, and Vietnam.

Code information

OTW2897¿ OTW3273¿ OTW3333¿ OTW3379¿ OTW3418¿ OTW3543¿ OTW3564¿ OTW3620¿ OTW3700¿ OTW3740¿ OTW3758¿ OTW3805¿ OTW3865¿ OTW3879¿ OTW3892¿ OTW3966¿ OTW4009¿ OTW4010¿ OTW4017¿ OTW4018¿ OTW4024¿ OTW4060¿ OTW4063¿ OTW4064¿ OTW4065¿ OTW4117¿ OTW4201¿ OTW4238¿ OTW4239¿ OTW2957¿ OTW3001¿ OTW3266¿ OTW3395¿ OTW3565¿ OTW3637¿ OTW3638¿ OTW3658¿ OTW3665¿ OTW3670¿ OTW3671¿ OTW3682¿ OTW3683¿ OTW3694¿ OTW3695¿ OTW3701¿ OTW3708¿ OTW3727¿ OTW3728¿ OTW3747¿ OTW3769¿ OTW3770¿ OTW3775¿ OTW3776¿ OTW3779¿ OTW3780¿ OTW3810¿ OTW3820¿ OTW3821¿ OTW3853¿ OTW3854¿ OTW3880¿ OTW3881¿ OTW3886¿ OTW3893¿ OTW3898¿ OTW3899¿ OTW3915¿ OTW3916¿ OTW3924¿ OTW3930¿ OTW3957¿ OTW3958¿ OTW3981¿ OTW3982¿ OTW4011¿ OTW4025¿ OTW4026¿ OTW4030¿ OTW4031¿ OTW4052¿ OTW4053¿ OTW4054¿ OTW4066¿ OTW4067¿ OTW4068¿ OTW4073¿ OTW4074¿ OTW4075¿ OTW4092¿ OTW4093¿ OTW4098¿ OTW4099¿ OTW4105¿ OTW4106¿ OTW4107¿ OTW4119¿ OTW4120¿ OTW4157¿ OTW4158¿ OTW4188¿ OTW4197¿ OTW4220¿ OTW4221¿ OTW4222¿ OTW4229¿ OTW4230¿ OTW4247¿ OTW4249¿ OTW4254¿ OTW4255¿ OTW4258¿ OTW4259¿ OTW4260¿ OTW4261¿ OTW4263¿ OTW4291¿ OTW4294¿ OTW4339¿ OTW4341¿ OTW4363¿ OTW4364¿ OTW2942¿ OTW2959¿ OTW2987¿ OTW3207¿ OTW3292¿ OTW3303¿ OTW3312¿ OTW3322¿ OTW3323¿ OTW3351¿ OTW3367¿ OTW3438A¿ OTW3511¿ OTW3635¿ OTW3636¿ OTW3660¿ OTW3661¿ OTW3667¿ OTW3688¿ OTW3689¿ OTW3706¿ OTW3707¿ OTW3711¿ OTW3717¿ OTW3718¿ OTW3734¿ OTW3750¿ OTW3783¿ OTW3784¿ OTW3794¿ OTW3795¿ OTW3827¿ OTW3828¿ OTW3851¿ OTW3852¿ OTW3862¿ OTW3863¿ OTW3882¿ OTW3896¿ OTW3897¿ OTW3907¿ OTW3908¿ OTW3931¿ OTW3936¿ OTW3944¿ OTW3945¿ OTW3962¿ OTW3963¿ OTW3978¿ OTW3979¿ OTW3995¿ OTW3996¿ OTW4033¿ OTW4034¿ OTW4057¿ OTW4058¿ OTW4084¿ OTW4085¿ OTW4127¿ OTW4128¿ OTW4140¿ OTW4141¿ OTW4146¿ OTW4147¿ OTW4176¿ OTW4177¿ OTW4183¿ OTW4184¿ OTW4191¿ OTW4192¿ OTW4193¿ OTW4232¿ OTW4317¿ OTW4336¿ OTW2872¿ OTW3191¿ OTW3439¿ OTW3537¿ OTW3623¿ OTW3662¿ OTW3668¿ OTW3669¿ OTW3690¿ OTW3691¿ OTW3712¿ OTW3735¿ OTW3766¿ OTW3771¿ OTW3812¿ OTW3822¿ OTW3823¿ OTW3836¿ OTW3849¿ OTW3850¿ OTW3874¿ OTW3888¿ OTW3903¿ OTW3925¿ OTW3932¿ OTW3946¿ OTW3947¿ OTW3968¿ OTW3969¿ OTW3987¿ OTW3988¿ OTW4020¿ OTW4021¿ OTW4036¿ OTW4037¿ OTW4038¿ OTW4079¿ OTW4080¿ OTW4087¿ OTW4196¿ OTW4267¿ OTW4268¿ OTW4269¿ OTW4270¿ OTW4275¿ OTW4276¿ OTW4277¿ OTW4288¿ ***Added 3/19/20*** OTW2847¿ OTW2863¿ OTW2871¿ ***Updated 4/14/20*** The firm noted upon review that some of the original recall lots were originally relabeled with different lot numbers for Non-US customers. Those additional lot numbers are as follow: OTW3461 OTW3501 OTW3591 OTW3628 OTW3737 OTW3742 OTW3753 OTW3813 OTW3859 OTW3861 OTW4208 ***Updated 8/21/20*** Lots OTW4180, OTW4187, OTW4198, OTW4202, OTW4218, OTW4246

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1567-2020
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for LeMaitre Vascular, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1567-2020. Report a data issue: hello@regdatalab.com.