Z-1569-2013 Class II Terminated

Recalled by Terumo Cardiovascular Systems Corporation — Ann Arbor, MI

FDA device recall Z-1569-2013 was initiated by Terumo Cardiovascular Systems Corporation on May 30, 2013 and is designated Class II. Reason for recall: Internal testing found certain Sarns" TCM II systems were released with incorrectly assembled heatsink assemblies which could cause the units to fail. The recall status is terminated (terminated Sep 17, 2013). Affected quantity: 7 units.

Recall details

Product Type
Devices
Report Date
Jul 3, 2013
Initiation Date
May 30, 2013
Termination Date
Sep 17, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
7 units

Product description

Sarns " TCM II Temperature Control Module The Sarns" TCM II Temperature Control Module is a source of temperature-controlled water for blood heat exchangers used in an extracorporeal circuit and is indicated for controlling and monitoring patient temperature

Reason for recall

Internal testing found certain Sarns" TCM II systems were released with incorrectly assembled heatsink assemblies which could cause the units to fail.

Distribution pattern

Worldwide Distribution including the countries of UAE (United Arab Emirates) HONG KONG and THAILAND. Not distributed in USA

Code information

Catalog Number 4416 and Serial Numbers: 4656, 4657, 4658,4659, 4661, 4662, and 4663.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1569-2013
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Terumo Cardiovascular Systems Corporation? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Terumo Cardiovascular Systems Corporation: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Terumo Cardiovascular Systems… →

Source: openFDA record for Z-1569-2013. Report a data issue: hello@regdatalab.com.