Z-1569-2025 Class II Ongoing

Recalled by Conformis Inc. — Wilmington, MA

Recall Details

Product Type
Devices
Report Date
April 23, 2025
Initiation Date
March 25, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
16

Product Description

restor3d Cordera Hip System. ACETABULAR LINER GRP C 36MM LIP. Model Number: HDL-060-C22L-000101 Acetabular Liner for Cordera Hip System

Reason for Recall

The hip cup liner that was packaged with the incorrect labeling. HDL-060-C22S-000101 from Lot 1788258 contained a hooded liner, but it was labeled to have a standard liner.

Distribution Pattern

US Nationwide distribution in the states of CA, FL, NV, TN.

Code Information

Model Number: HDL-060-C22L-000101 UDI-DI: M572HDL060C22L011. Lot: 1788258