Z-1574-2023 Class II Ongoing

Recalled by Medtronic Sofamor Danek USA Inc — Memphis, TN

FDA device recall Z-1574-2023 was initiated by Medtronic Sofamor Danek USA Inc on Nov 19, 2021 and is designated Class II. Reason for recall: Two lots of the 5598007 Distractor Rack which were not manufactured to specification. The crest width of the gear is too wide and may cause difficulty or the inability to operate the Distractor. The recall status is ongoing. Affected quantity: 4 units.

Recall details

Product Type
Devices
Report Date
May 17, 2023
Initiation Date
Nov 19, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4 units

Product description

Distractor Rack, REF 5598007, which may be packaged in Modular Distractor Set, Loaner Kit Number SPS03139; for use in spinal surgery

Reason for recall

Two lots of the 5598007 Distractor Rack which were not manufactured to specification. The crest width of the gear is too wide and may cause difficulty or the inability to operate the Distractor.

Distribution pattern

US: IN, CT, NE

Code information

UDI/DI (GTIN) 00763000401535, Lot Numbers EM21D006, EM21D007; Loaner Kit Lot numbers: 0072, 0073, 0077, 0098, 0103

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1574-2023
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Medtronic Sofamor Danek USA Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Medtronic Sofamor Danek USA Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Medtronic Sofamor Danek USA Inc →

Source: openFDA record for Z-1574-2023. Report a data issue: hello@regdatalab.com.