Z-1574-2023 Class II Ongoing

Recalled by Medtronic Sofamor Danek USA Inc — Memphis, TN

Recall Details

Product Type
Devices
Report Date
May 17, 2023
Initiation Date
November 19, 2021
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4 units

Product Description

Distractor Rack, REF 5598007, which may be packaged in Modular Distractor Set, Loaner Kit Number SPS03139; for use in spinal surgery

Reason for Recall

Two lots of the 5598007 Distractor Rack which were not manufactured to specification. The crest width of the gear is too wide and may cause difficulty or the inability to operate the Distractor.

Distribution Pattern

US: IN, CT, NE

Code Information

UDI/DI (GTIN) 00763000401535, Lot Numbers EM21D006, EM21D007; Loaner Kit Lot numbers: 0072, 0073, 0077, 0098, 0103