Z-1588-2026 Class I Ongoing

Recalled by Navajo Manufacturing Company — Denver, CO

Recall Details

Product Type
Devices
Report Date
April 1, 2026
Initiation Date
February 18, 2026
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
28,457

Product Description

Handy Solutions Neck & Shoulders Heating Pad, Number: 25607, Manufactured by Ninbo Unico Products Co., Ltd

Reason for Recall

If heating pad is folded while in use, the power density increases in the folded area causing excessive temperatures, which can also be caused by placing pad under the body, then the high limit device is unable to sense the higher temperature so power is not terminated. May cause a burning smell prior to melting/burning potentially causing burns to persons and property.

Distribution Pattern

US Nationwide distribution in the states of CA, MS, FL, IN, LA, VA, TX.

Code Information

UPC: 0-24291-25607-6, UDI-DI: 00024291256076