Z-1588-2026 Class I Ongoing
FDA device recall Z-1588-2026 was initiated by Navajo Manufacturing Company on Feb 18, 2026 and is designated Class I. Reason for recall: If heating pad is folded while in use, the power density increases in the folded area causing excessive temperatures, which can also be caused by placing pad under the body, then the high limit device… The recall status is ongoing. Affected quantity: 28,457.
Recall details
- Product Type
- Devices
- Report Date
- Apr 1, 2026
- Initiation Date
- Feb 18, 2026
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 28,457
Product description
Reason for recall
Distribution pattern
US Nationwide distribution in the states of CA, MS, FL, IN, LA, VA, TX.
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1588-2026
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026