Z-1592-2013 Class II Terminated

Recalled by Medtronic Sofamor Danek USA Inc — Memphis, TN

FDA device recall Z-1592-2013 was initiated by Medtronic Sofamor Danek USA Inc on May 28, 2013 and is designated Class II. Reason for recall: Kits contain a collagen sponge recalled by Integra due to possible elevated endotoxin levels resulting from a process deviation. The recall status is terminated (terminated May 13, 2014). Affected quantity: 16994 units.

Recall details

Product Type
Devices
Report Date
Jul 3, 2013
Initiation Date
May 28, 2013
Termination Date
May 13, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
16994 units

Product description

INFUSE(R) Bone Graft,LARGE II KIT, REF 7510800, Medtronic Sofamore Danek USA, Inc., 1800 Pyramid Place, Memphis, Tennessee 38132 orthopaedic

Reason for recall

Kits contain a collagen sponge recalled by Integra due to possible elevated endotoxin levels resulting from a process deviation.

Distribution pattern

Worldwide distribution: USA (nationwide including Puerto Rico) and Internationally to Australia, Brazil, Canada, Chile, Hong Kong, India, Mexico, Netherlands, , New South Wales, Panama, and Paraguay.

Code information

Lots: M111103AAG, M111052AAQ, M111052AAR, M111052AAP, M111052AAL, M111059AAC, M111059AAF, M111052AAF, M111059AAH, M111103AAB, M111059AAG, M111052AAK, M111052AAJ, M111059AAB, M111059AAK, M111052AAI, M111052AAO, M111052AAH, M111052AAG, M111059AAD

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1592-2013
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Medtronic Sofamor Danek USA Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Medtronic Sofamor Danek USA Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Medtronic Sofamor Danek USA Inc →

Source: openFDA record for Z-1592-2013. Report a data issue: hello@regdatalab.com.