Z-1595-2024 Class II Ongoing

Recalled by Raye's Inc. — Hays, KS

Recall Details

Product Type
Devices
Report Date
April 24, 2024
Initiation Date
January 25, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,149 units

Product Description

SW Bari Lift & Transfer, Model Number 38060000

Reason for Recall

Retrospective engineering analysis against current ISO standards for hoists has determined: 1. The device cannot physically perform at maximum documented weight specifications when a factor of safety of two times the maximum load is applied. 2. The device is not equipped with a safety device to ensure a person with disability would not fall in the event of a single-fault condition of the lifting device.

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Australia, UK.

Code Information

UDI/DI 00845699001040, All serial numbers