Z-1595-2024 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- April 24, 2024
- Initiation Date
- January 25, 2024
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1,149 units
Product Description
SW Bari Lift & Transfer, Model Number 38060000
Reason for Recall
Retrospective engineering analysis against current ISO standards for hoists has determined: 1. The device cannot physically perform at maximum documented weight specifications when a factor of safety of two times the maximum load is applied. 2. The device is not equipped with a safety device to ensure a person with disability would not fall in the event of a single-fault condition of the lifting device.
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Australia, UK.
Code Information
UDI/DI 00845699001040, All serial numbers