Z-1595-2024 Class II Ongoing

Recalled by Raye's Inc. — Hays, KS

FDA device recall Z-1595-2024 was initiated by Raye's Inc. on Jan 25, 2024 and is designated Class II. Reason for recall: Retrospective engineering analysis against current ISO standards for hoists has determined: 1. The device cannot physically perform at maximum documented weight specifications when a factor of safety… The recall status is ongoing. Affected quantity: 1,149 units.

Recall details

Product Type
Devices
Report Date
Apr 24, 2024
Initiation Date
Jan 25, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,149 units

Product description

SW Bari Lift & Transfer, Model Number 38060000

Reason for recall

Retrospective engineering analysis against current ISO standards for hoists has determined: 1. The device cannot physically perform at maximum documented weight specifications when a factor of safety of two times the maximum load is applied. 2. The device is not equipped with a safety device to ensure a person with disability would not fall in the event of a single-fault condition of the lifting device.

Distribution pattern

Worldwide distribution - US Nationwide and the countries of Australia, UK.

Code information

UDI/DI 00845699001040, All serial numbers

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1595-2024
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Raye's Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1595-2024. Report a data issue: hello@regdatalab.com.