Z-1613-2021 Class II Ongoing
FDA device recall Z-1613-2021 was initiated by Medtronic Inc. on Feb 26, 2021 and is designated Class II. Reason for recall: Use of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor glucose readings. The recall status is ongoing. Affected quantity: 15,162 units.
Recall details
- Product Type
- Devices
- Report Date
- May 26, 2021
- Initiation Date
- Feb 26, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 15,162 units
Product description
Reason for recall
Distribution pattern
Worldwide distribution.
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1613-2021
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026