Z-1616-2014 Class II Terminated

Recalled by Roche Diagnostics Operations, Inc. — Indianapolis, IN

FDA device recall Z-1616-2014 was initiated by Roche Diagnostics Operations, Inc. on Apr 14, 2014 and is designated Class II. Reason for recall: Roche Diabetes Care has become aware the ACCU-CHEK Compact Plus test strips may produce erroneously low blood glucose readings in patients undergoing Ceftriaxone therapy. The interference is not… The recall status is terminated (terminated Sep 18, 2014). Affected quantity: 590,136 boxes.

Recall Details

Product Type
Devices
Report Date
May 21, 2014
Initiation Date
Apr 14, 2014
Termination Date
Sep 18, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
590,136 boxes

Product Description

ACCU-CHEK Compact Plus Test Strips For use with: ACCU-CHEK Compact Plus Meters; 17 count sample strip 51 count retail strip; 102 count retail strip; 51 count MedBen strip 51 count Mail Order strip; Sample Kit 16 per case. The ACCU-CHEK Compact Plus Test Strips are for use with the ACCU-CHEK Compact Plus Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips or palm.

Reason for Recall

Roche Diabetes Care has become aware the ACCU-CHEK Compact Plus test strips may produce erroneously low blood glucose readings in patients undergoing Ceftriaxone therapy. The interference is not described in the product labeling.

Distribution Pattern

Nationwide Distribution.

Code Information

5599423160, 5599415160, 5919967160, 5919959160, 5919983160, 5907675001

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)