Z-1622-2013 Class II Terminated

Recalled by Maquet Cardiovascular Us Sales, Llc — Wayne, NJ

FDA device recall Z-1622-2013 was initiated by Maquet Cardiovascular Us Sales, Llc on Apr 26, 2013 and is designated Class II. Reason for recall: MAQUET has identified a potential issue with the extension device handle (part number 1433.62A1) and will make enhancements to the YUNO table (part number 1433.01F0). Identification of the… The recall status is terminated (terminated Apr 29, 2014). Affected quantity: 13 Tables; 24 Extension devices.

Recall details

Product Type
Devices
Report Date
Jul 10, 2013
Initiation Date
Apr 26, 2013
Termination Date
Apr 29, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
13 Tables; 24 Extension devices

Product description

MAQUET Getinge Group Mobile Operating Table YUNO OTN GA 1433.01 EN 05. The YUNO OTN is used to support and position patients for surgical procedures immediately before, during and after the surgical phase. The table top is radio translucent and enables intraoperative use of X-ray equipment.

Reason for recall

MAQUET has identified a potential issue with the extension device handle (part number 1433.62A1) and will make enhancements to the YUNO table (part number 1433.01F0). Identification of the Extension device issue: By fully actuating the lever of the extension device handle, the operator's skin can enter a gap between the handle lever and a recess in the handle bar. When the lever is released,

Distribution pattern

Worldwide Distribution - USA including Arizona, Georgia, North Dakota, South Carolina, Texas and Tennessee, Internationally to United Arab Emirates, Austria, Brazil, Canada, Switzerland, China, Germany, Spain, France, United Kingdom, Italy, Japan, Singapore, Thailand, Turkey, Serbia, and South Africa.

Code information

YUNO OTN Mobile Operating Table: 1433.01F0 Extension Device: 1433.62A1

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1622-2013
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1622-2013. Report a data issue: hello@regdatalab.com.