Z-1627-2021 Class II Terminated

Recalled by Philips Respironics, Inc. — Murrysville, PA

FDA device recall Z-1627-2021 was initiated by Philips Respironics, Inc. on Apr 6, 2021 and is designated Class II. Reason for recall: Certain devices were built with motor assemblies that could contain non-conforming plastic material. This could result in 1. Exposure to off-gassing of Volatile Organic Compounds not normally present… The recall status is terminated (terminated Jan 17, 2024). Affected quantity: 1 unit.

Recall details

Product Type
Devices
Report Date
Jun 2, 2021
Initiation Date
Apr 6, 2021
Termination Date
Jan 17, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1 unit

Product description

BiPAP A40 System Silver, JP: Continuous ventilator, non-life-supporting. Material Number: R1111177

Reason for recall

Certain devices were built with motor assemblies that could contain non-conforming plastic material. This could result in 1. Exposure to off-gassing of Volatile Organic Compounds not normally present creating a potential biosafety or toxicological hazard and 2. Sudden failure of the device causing a Ventilator Inoperative condition with the potential for asphyxia if not immediately identified and addressed by the care provider.

Distribution pattern

Distribution US nationwide to states of: LA, PA, WI, CO, WV, WA, CA, TX, MO, OK, FL, and TN. Brazil, China, and Japan

Code information

Serial Number: V30562261B40E

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1627-2021
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Philips Respironics, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Philips Respironics, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Philips Respironics, Inc. →

Source: openFDA record for Z-1627-2021. Report a data issue: hello@regdatalab.com.