Z-1629-2026 Class II Ongoing

Recalled by Orthalign, Inc — Aliso Viejo, CA

Recall Details

Product Type
Devices
Report Date
April 1, 2026
Initiation Date
February 10, 2026
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
7 units

Product Description

Drill Plate, Ortho Dev. Trimax - Pegged, PREF Size 1 -5 REF:402566 component Lantern Knee System

Reason for Recall

Drill plates manufactured out of specification can result in a bone cut in excess 00800"(2.8448mm)

Distribution Pattern

US Nationwide distribution in the state of UT.

Code Information

Lot # 25091103BB/ UDI: 00810832032427