Z-1629-2026 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- April 1, 2026
- Initiation Date
- February 10, 2026
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 7 units
Product Description
Drill Plate, Ortho Dev. Trimax - Pegged, PREF Size 1 -5 REF:402566 component Lantern Knee System
Reason for Recall
Drill plates manufactured out of specification can result in a bone cut in excess 00800"(2.8448mm)
Distribution Pattern
US Nationwide distribution in the state of UT.
Code Information
Lot # 25091103BB/ UDI: 00810832032427