Z-1633-2015 Class II Terminated

Recalled by Stryker Howmedica Osteonics Corp. — Mahwah, NJ

FDA device recall Z-1633-2015 was initiated by Stryker Howmedica Osteonics Corp. on Aug 15, 2014 and is designated Class II. Reason for recall: The peel strength of the inner blister may have been below internal validated requirements. The strength of the seal is directly related to long term shelf life of the sterile barrier. The recall status is terminated (terminated Jan 5, 2018). Affected quantity: 1,147 in total.

Recall details

Product Type
Devices
Report Date
May 27, 2015
Initiation Date
Aug 15, 2014
Termination Date
Jan 5, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,147 in total

Product description

Stryker Howmedica Osteonics RESTORATION Acetabular Augment System; Howmedica Osteonics Corp., A subsidiary of Stryker Corporation Made in USA Sterile. Hip prosthesis component. The RESTORATION Acetabular Wedge Augments provide multiple options to address the wide range of bone deficiencies encountered in acetabular revision. The augments provide support for the shell in the acetabulum with superior and/or posterior defects.

Reason for recall

The peel strength of the inner blister may have been below internal validated requirements. The strength of the seal is directly related to long term shelf life of the sterile barrier.

Distribution pattern

Worldwide Distribution: US (nationwide) including PR and countries of: Australia, Canada, China, Brazil, France, Hong Kong, India, Italy, Mexico, Netherlands, Singapore, South Africa, Sweden, and Spain.

Code information

Ref #:Lot Number - 5096-4615:MMR8M8; 5096-4615:MMRLYD; 5096-4615:MNAYT7; 5096-5015:MMPO6K; 5096-5015:MMRWOJ; 5096-5015:MNADVA; 5096-5815:MNA40L; 5096-5820:MMR6HX; 5096-5825:MMRLD7; 5096-6225:MMPWO5; 5096-6225:MMRM2Y; 5096-6625:MMNM07

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1633-2015
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1633-2015. Report a data issue: hello@regdatalab.com.