Z-1634-2015 Class II Terminated

Recalled by Stryker Howmedica Osteonics Corp. — Mahwah, NJ

FDA device recall Z-1634-2015 was initiated by Stryker Howmedica Osteonics Corp. on Aug 15, 2014 and is designated Class II. Reason for recall: The peel strength of the inner blister may have been below internal validated requirements. The strength of the seal is directly related to long term shelf life of the sterile barrier. The recall status is terminated (terminated Jan 5, 2018). Affected quantity: 1,147 in total.

Recall details

Product Type
Devices
Report Date
May 27, 2015
Initiation Date
Aug 15, 2014
Termination Date
Jan 5, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,147 in total

Product description

Stryker Orthopaedics Scorpio femoral component. Howmedica Osteonics Corp., A subsidiary of Stryker Corporation Made in USA Sterile. For use in total reconstructive replacement of the knee joint.

Reason for recall

The peel strength of the inner blister may have been below internal validated requirements. The strength of the seal is directly related to long term shelf life of the sterile barrier.

Distribution pattern

Worldwide Distribution: US (nationwide) including PR and countries of: Australia, Canada, China, Brazil, France, Hong Kong, India, Italy, Mexico, Netherlands, Singapore, South Africa, Sweden, and Spain.

Code information

Item number 26-3005L Lot numbers MNH0ND, MNH3V7, MNH8XR, MNHN0E, MNH6Y5; 70-3003L Lot MNE6NM; 70-3003R Lot MNE963; 70-3007L Lots MNE58K, MNE6HL, MNE8MN, MNEKXP; 70-3007R Lots MNEV2Y, MNEVDV, MNEN3D, MNEX43; 70-3013L Lot MMMXTY; 70-4104L Lots MNE0Y8, MNE2J2, MND743, MNDVDT; 71-3009L Lots MNE12D, MNEALW, MNEEE4, MNEXLK, MNEXV0; 71-3009R Lots MNELHM, MNEYNE; 71-3013L Lot MMR24J; 71-4104L Lot MMAKXS; 71-4503R Lots MMT7P8, MMT8J3, MNDKJ7; 71-4504L Lots MNE5D4, MNEHV1; 71-4504R Lots MND272, MNDD9D, MNEL8W; 71-4505L Lots MMT4KY, MMTPHS, MNAJ78, MNAN2L, MNDP2N; 71-4505R Lots MMTEW1, MNA062, MNANVA; 71-4506L Lots MNAJ1W, MNARAR, MNEL1Y, MNEW53; 71-4507L Lots MMT971, MMT9R5X1, MMTJRT, MNALJA, MND1V8; 71-4507R Lots MNA17E, MND4WP, MNEAAD, MNEM55; 71-4509R Lots MMR2HL, MND9AE; 80-4403L Lots MNE0LE, MNEJ70, MNEK3L; 80-4404R Lot MNHTRK, 80-4405L Lot MNERD1; 80-4405R Lots MNEEND, MNEHNO, MNEK42, MNEMNR, MNEMXH, MNEY4A; 80-4406L Lots MNA4HR, MNEH40, MNE37L, MNE769, MNE9RD, MNEATP, MNEX06, MNEXA8; 80-4407R Lots MNELAL, MNEP9K, MNEPVE; 80-4408L Lot MNEX9T; 80-4409L Lots MNE7H9, MNEDWK, MNEDXN; 80-4409R Lots MNE0R2, MNETXP, MNEVTV; 80-4411L Lots MNEN81, MNEWN4; 80-6404R Lot MNDKD0; 80-6405R Lots MMTA8L, MNA175; 80-6407L Lot MMRV9D; 80-6411L Lot MMP7TS; 81-4404R Lots MNE34H, MNE9WN, MNEKY7; 81-4405L Lots MNE3XK, MNE4OR, MNE473, MNE50T, MNE6A5, MNE7NN, MNE9DD, MNEDX6, MNEJ0N, MNELK5, MNERR2; 81-4405R Lots MNE5PE, MNE5W5, MNEA1T, MNEH09, MNEK9M, MNEP17, MNEPET, MNEPX6, MNER3V, MNEWW7, MNEXDP, MNEXLL, MNEY3R; 81-4406L Lots MNE0J3, MNE5D6, MNEA42, MNEHL1, MNEM0H, MNEMD4, MNEMJ8, MNEMPD, MNENAR, MNEPN6, MNEW72; 81-4406R Lots MNE1XK, MNE5D2, MNEA75, MNED3A, MNEVLY, MNEYHH; 81-4407L Lots MNE127, MNE26L, MNE3KJ, MNE63T, MNE6XK, MNE71T, MNE8PT, MNE996, MNEPYY; 81-4407R Lots MMPE33, MNEOJ9, MNE217, MNE2NH, MNE3MH, MNE525, MNE5YP, MNE6WT, MNE7V2, MNE9J1, MNEA62, MNEAR2, MNEAXJ, MNED83, MNENNE, MNELEP, MNENM4, MNENY7, MNEPEP, MNERAV, MNETYX, MNEVN7, MNEW3Y, MNEW56, MNEWN0, MNEX5E, MNEXV1, MNEXVT, MNEY27, MNEYR6, MNHTW5; 81-4408L Lots MNE2VE, MNEAMY, MNEEM0, MNEYRT, MNEYYK; 81-4408R Lots MNEO89, MNE2A1, MNE4DK, MNE7XV, MNE8N4, MNE9M9, MNEA4T, MNEAJ8, MNEMLA, MNET9J; 81-4409L Lots MNE4W5, MNE93H, MNEEEN, MNEVOY, MNEY57, MNEY74, MNE159, MNE1H2, MNE2DK, MNEM2D, MNENLJ, MNER77, MNEV0J; 81-4411R Lot MMTVXW; 81-4413L Lot MMP0MP; 81-6404R Lot MMTR58; 81-6407L Lot MMTDN7.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1634-2015
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1634-2015. Report a data issue: hello@regdatalab.com.