Z-1637-2016 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- May 18, 2016
- Initiation Date
- February 2, 2015
- Termination Date
- January 10, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 3,760 units
Product Description
Axial Torque Limiting Driver (1st generation) Product Usage: The Bradshaw Medical Axial Torque Limiting Driver is a reusable surgical instrument designed to limit the magnitude of applied (input) torque transferred to an external (output) device or fastener in the clockwise direction. The device is a non-adjustable mechanical click style hand torque tool which presents an audible and tactile click when the maximum desired torque is reached.
Reason for Recall
A component failure could cause a situation where the applied torque was significantly higher than the calibrated torque setting.
Distribution Pattern
US Nationwide Distribution in the states of: CA, FL, IN, MA, MI, OH, PA, TN UT, VA and the country of UNITED KINGDOM
Code Information
Part Number BMI # Customer Lot Code 11-5000-018 512833 102551 512835 102848 527192 OP-CS-I-0026 TO 0046 550398 OP-CS-I-0223 TO 0232 561041 561041-001 TO -030 6200-1109 513884 2009300543 514227 2009300643 514228 2009300743 514229 2009300843 514230 2009300993 516828 2009411377 518909 2010082257 519942 2010152571 522701 2010373025-2010373174 14-525035-00 538133 538133 001-030 538134 538134 001-015 548080 548080 001-015 550475 550475 001-015 552169 552169 001-020 553131 553131 001-015 553710 553710 001-015 558873 558873 001-010 567568 567568 001-010 569141 569141 001-015 PARD01085 530249 351400 001-002 L070-0045 570659 1525-10-1 THRU 1525-10-2 694.002 523947 BRL237AA001 523948 BRL237AA 002 TO 092 527249 BRM032AA-001 TO 015 529395 BRM061AA 001 TO 030 528469 BRM061AA 001 TO 050 534799 BRM259FA 535830 BRM292GA 537840 BRM362AA 539698 BRN046AA 540726 BRN073AA 543837 BRN171BA 544323 BRN194AA003 544323-1 BRN194AA 001-002 AND 004-014 547606 BRN285EA 548612 BRN317DA 549604 BRN348VA 554050 BRP102AA 555027 BRP107CA-001 TO 010 CHANGED TO BRP214AA AFTER REWORK 556873 BRP192VA 558042 BRP225AA 560135 BRP283AA 561731 BRP324AA 562453 BRP345AA 563429 BRR010FA 565115 BRR048AA 571756 BRR255BA -001 TO 010 572999 BRR293FA ALSO 572999-001 TO 025 K2-01-2473-01 503404 LDS-001d to -015 509497 AAHG-001 TO -030 509498 AAHV-001 TO -025 514731 ACAV-001 TO -030 518702 AEXJ-001 TO -035 520216 AFXM-001 TO -010 522688 AGRK-001 TO -085 524265 AJNC-001 TO -015 526562 AKLA-001 TO -050 532140 ATDE-001 TO -015 533737 AUAF-001 TO -020 537141 AYGY-001 TO -015 541350 BDRY-001 TO -025 549451 BMCU-001 TO -019 550756 BMCU-020 TO -070 552289 BRRL-001 TO -030 556336 BWEE-001 TO -015 563714 CHCU-001 TO -015 565297 CLWD-001 TO -045 K2-08-2294-01 516879 ADRF-001 TO -010 521965 AGKL-001 TO -025 521966 AGKL-026 TO -040 525435 AJYW-001 TO -010 525855 AJYW-001 TO -040 527600 AKYY-010 TO -030 532139 ATCU-001 TO -035 541358 BDVV-001 TO -015 552078 BRKB-001 TO -060 552286 BRRR-001 TO -060 575988 DHDB-001 to -020 K2-12-2044-01 509002 AAEM-001 TO -006 518290 AEEJ-010 TO -020 523307 AJAK-001 TO -015 529220 AMEM-001 TO -010 554653 BVLJ-001 TO -010 557778 BYCR-001 TO -015 565304 CLYJ-001 TO -005 570207 DBDH-001 TO -010 11-K03-5001 566401 B411 TO B420 281-1-10220 535841 535841-001 to 005 535842 535842-001 to 031 541403 541403 005, 006, 030 541403-1 541403-1 001-064 543008 543008-001 to 030 544168 544168-001 to 010 544169 544169-001 to 015 545627 545627-001 to 005 546471 546471-001 to -010 550013 550013-001 to 020 550831 550831-001 to 010 552163 552163-001 to 005 554468 554468-001 to 010 558407 558407-001 to 010 560471 560471-001 to 015 563037 563037-001 to 025 571303 571303-001 to 005 652-1-10300 500923 46420 510872 510872 512031 512031 515511 515511 525260 525260 563264 563264-01 to 05 723-454 566247 566247-001 to 003 91-1112 512367 AL512367B 001 TO 008 520386 AL520386B 001 TO 012 521170 AL521170B 001 TO 003 533351 AL533351B 001-015 545459 AL545459B-001 TO 015 558280 AL558280B 001 TO 010 560756 AL560756B 001 TO 010 567356 AL567356B-001 TO 010 95-0127 500382 AL500382A 504238 AL504238A 001 TO 045 507588 AL507588A 001 TO 015 507589 AL507589A 001 TO 010 515870 AL515870A 001 TO 008 520385 AL520385A 001 TO 020 560882 AL560882A-001 TO 010 563933 AL563933A-001 TO 010 572915 AL572915A-001 TO 015 03.811.109 544776 113056-01 562281 7532121-01 TO 25 15-614-000-XBOLT 557962 557962-001 TO 010 565491 565491-001 TO 025 908.133-1 535886 1011-6 559488 0913-6 568816 0214-6 570566 1014-6 X060-0323 522672 522672 525941 525941 535244 535244 535245 535245 535817 535817 525942 525942 531776 53177