Z-1646-2019 Class II Terminated

Recalled by Maquet Cardiovascular, LLC — Wayne, NJ

FDA device recall Z-1646-2019 was initiated by Maquet Cardiovascular, LLC on Apr 18, 2019 and is designated Class II. Reason for recall: Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path… The recall status is terminated (terminated Jul 28, 2025).

Recall details

Product Type
Devices
Report Date
Jun 5, 2019
Initiation Date
Apr 18, 2019
Termination Date
Jul 28, 2025
Voluntary/Mandated
Voluntary: Firm initiated

Product description

Maquet Getinge: (1) Material: 701053941R01 BEQ-TOP 9005 3/8 Custom Pk w/RF32 & HMOD (2) Material: 701053941R02- BEQ-TOP 9005 3/8 Custom Pk w RF32 & HMOD Material:

Reason for recall

Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured

Distribution pattern

Nationwide

Code information

Batch Numbers: (1) 3000051652 3000054808 3000058733 3000062473 3000071201 3000073170 3000075123 (2) 3000076148 3000081600

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1646-2019
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1646-2019. Report a data issue: hello@regdatalab.com.