Z-1663-2019 Class II Terminated

Recalled by Maquet Cardiovascular, LLC — Wayne, NJ

FDA device recall Z-1663-2019 was initiated by Maquet Cardiovascular, LLC on Apr 18, 2019 and is designated Class II. Reason for recall: Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path… The recall status is terminated (terminated Jul 28, 2025).

Recall details

Product Type
Devices
Report Date
Jun 5, 2019
Initiation Date
Apr 18, 2019
Termination Date
Jul 28, 2025
Voluntary/Mandated
Voluntary: Firm initiated

Product description

Maquet Getinge- (1)Material: 701065765R01 BO-TOP 36504 Adult ECC Pack (2)Material: 701065765R02 BO-TOP 36504 Adult ECC Pack

Reason for recall

Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured

Distribution pattern

Nationwide

Code information

Batch Numbers: (1)3000058184, 3000067382, 3000069380, 3000071967, 3000074322 (2)3000077734

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1663-2019
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1663-2019. Report a data issue: hello@regdatalab.com.