Z-1664-2021 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- May 26, 2021
- Initiation Date
- April 7, 2021
- Termination Date
- June 1, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- N/A
Product Description
REF 69655405, Groshong NXT ClearVue Catheter with Sherlock Tip Location System (TLS) Stylet, Basic Kit, 4F, Single-Lumen, TLS, UDI: (01)10801741147521 - Product Usage: is designed for use when central venous catheterization is prescribed.
Reason for Recall
Catheters contain two issues: 1. the inner diameter of the distal connector may be reduced resulting in inability to pass the catheter through the connector and 2. the catch slot on the distal connecter may be deformed resulting in the inability to properly lock
Distribution Pattern
Worldwide distribution - US Nationwide distribution in the states of AK, AL, CA, CO, FL, GA, IA, IL, IN, KS, KY, LA, MA, MN, MO, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI and the countries of Australia, Belgium, Canada, China, Japan, Taiwan, Hong Kong, Malaysia, Chile, Bangladesh, Korea, Brazil, India, Singapore, Thailand.
Code Information
Lot No. redx4687, rees1313, reev1270