Z-1665-2021 Class II Terminated

Recalled by Bard Access Systems Inc. — Salt Lake City, UT

Recall Details

Product Type
Devices
Report Date
May 26, 2021
Initiation Date
April 7, 2021
Termination Date
June 1, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
N/A

Product Description

REF 91660417, Groshong NXT ClearVue Catheter with Sherlock 3CG Tip Positioning System (TPS) Stylet Tip, Basic Kit, 4F, Single-Lumen, TPS, CE 2797, UDI: (01)10801741107822 - Product Usage: is designed for use when central venous catheterization is prescribed.

Reason for Recall

Catheters contain two issues: 1. the inner diameter of the distal connector may be reduced resulting in inability to pass the catheter through the connector and 2. the catch slot on the distal connecter may be deformed resulting in the inability to properly lock

Distribution Pattern

Worldwide distribution - US Nationwide distribution in the states of AK, AL, CA, CO, FL, GA, IA, IL, IN, KS, KY, LA, MA, MN, MO, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI and the countries of Australia, Belgium, Canada, China, Japan, Taiwan, Hong Kong, Malaysia, Chile, Bangladesh, Korea, Brazil, India, Singapore, Thailand.

Code Information

Lot No. redx0702 redx1450 redx2338 redx4128 reep0823 reeq3588 reer1040 reer3480 rees0567 rees1246 rees2299 reew1849 reew4279 reex0353 reex3469