Z-1666-2024 Class I Ongoing

Recalled by Boston Scientific Corporation — Maple Grove, MN

FDA device recall Z-1666-2024 was initiated by Boston Scientific Corporation on Mar 21, 2024 and is designated Class I. Reason for recall: An increase in complaints related to an inability to advance the guidewire through the lumen of the device for certain batches. There is evidence of the polyurethane layer delaminating and in some… The recall status is ongoing. Affected quantity: 46860 catheters.

Recall Details

Product Type
Devices
Report Date
May 22, 2024
Initiation Date
Mar 21, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
46860 catheters

Product Description

Expo 5F Selective Angiographic Catheters, MODEL - 5F EXPO MULTIPACK (5PKX3), REF H749085263002; cardiac catheter

Reason for Recall

An increase in complaints related to an inability to advance the guidewire through the lumen of the device for certain batches. There is evidence of the polyurethane layer delaminating and in some instances material detachment in the catheters inner lining.

Distribution Pattern

Worldwide distribution. US, US Territories, EMEA, Canada, LATAM and Asia Pacific countries/regions.

Code Information

UDI/DI 08714729187868, Batch Numbers: 60427727, 60456253, 60462030, 60476271, 60500758, 60530055, 60436627, 60460188, 60474027, 60477271, 60502673, 60537578, 60438198, 60461419, 60475349, 60486289, 60515228

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1666-2024