Z-1672-2023 Class II Ongoing
Recalled by Smith & Nephew, Inc.
— Memphis, TN
FDA device recall Z-1672-2023 was initiated by Smith & Nephew, Inc. on May 2, 2023 and is designated Class II. Reason for recall: Recent complaint data indicates that the revision rate may be trending higher than corresponding similar devices in global joint replacement registries. The data identifies a potential signal that the… The recall status is ongoing. Affected quantity: 2146 units.
Recall details
- Product Type
- Devices
- Report Date
- Jun 7, 2023
- Initiation Date
- May 2, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2146 units
Product description
Smith&Nephew ENGAGE Cementless Partial Knee System, Coated Tibial Tray, Part Numbers: a) REF 1-10012-100, SIZE 1-LEFT MEDIAL; b) REF 1-10012-150, SIZE 1-RIGHT MEDIAL; c) REF 1-10012-200, SIZE 2-LEFT MEDIAL; d) REF 1-10012-250, SIZE 2-RIGHT MEDIAL; e) REF 1-10012-300, SIZE 3-LEFT MEDIAL; f) REF 1-10012-350, SIZE 3-RIGHT MEDIAL; g) REF 1-10012-400, SIZE 4-LEFT MEDIAL; h) REF 1-10012-450, SIZE 4-RIGHT MEDIAL; i) REF 1-10012-500, SIZE 5-LEFT MEDIAL; j) REF 1-10012-550, SIZE 5-RIGHT MEDIAL; k) REF 1-10012-600, SIZE 6-LEFT MEDIAL; l) REF 1-10012-650, SIZE 6-RIGHT MEDIAL; Unicondylar knee prothesis
Reason for recall
Recent complaint data indicates that the revision rate may be trending higher than corresponding similar devices in global joint replacement registries. The data identifies a potential signal that the performance is an outlier versus the state of the art with respect to the risk for revision.
Distribution pattern
US Nationwide distribution.
Code information
a) REF 1-10012-100, UDI/DI 00885556873113; b) REF 1-10012-150, UDI/DI 00885556873120; c) REF 1-10012-200, UDI/DI 00885556873137; d) REF 1-10012-250, UDI/DI 00885556873144; e) REF 1-10012-300, UDI/DI 00885556873151; f) REF 1-10012-350, UDI/DI 00885556873168; g) REF 1-10012-400, UDI/DI 00885556873175; h) REF 1-10012-450, UDI/DI 00885556873182; i) REF 1-10012-500, UDI/DI 00885556873199; j) REF 1-10012-550, UDI/DI 00885556873205; k) REF 1-10012-600, UDI/DI 00885556873212; l) REF 1-10012-650, UDI/DI 00885556873229; ALL BATCHES
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1672-2023
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Smith & Nephew, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Smith & Nephew, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about Smith & Nephew, Inc. →