Z-1679-2014 Class II Terminated
FDA device recall Z-1679-2014 was initiated by Cincinnati Sub-Zero Products Inc on Jan 8, 2014 and is designated Class II. Reason for recall: The incorrect voltage component was placed into the device during manufacturing. This may make the device unable to maintain patient temperature during cooling mode. The device will continue to… The recall status is terminated (terminated Jun 6, 2014). Affected quantity: 7 units.
Recall details
- Product Type
- Devices
- Report Date
- Jun 11, 2014
- Initiation Date
- Jan 8, 2014
- Termination Date
- Jun 6, 2014
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 7 units
Product description
Reason for recall
Distribution pattern
The product was distributed to one customer in MI, and internationally (Taiwan)