Z-1679-2014 Class II Terminated

Recalled by Cincinnati Sub-Zero Products Inc — Cincinnati, OH

FDA device recall Z-1679-2014 was initiated by Cincinnati Sub-Zero Products Inc on Jan 8, 2014 and is designated Class II. Reason for recall: The incorrect voltage component was placed into the device during manufacturing. This may make the device unable to maintain patient temperature during cooling mode. The device will continue to… The recall status is terminated (terminated Jun 6, 2014). Affected quantity: 7 units.

Recall details

Product Type
Devices
Report Date
Jun 11, 2014
Initiation Date
Jan 8, 2014
Termination Date
Jun 6, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
7 units

Product description

115 V Blanketrol II, Model 222S Hyper-Hypothermia System . Used to lower or to raise a patients temperature and /or maintain a desired patient temperature through conductive heat transfer. Part number 86165.

Reason for recall

The incorrect voltage component was placed into the device during manufacturing. This may make the device unable to maintain patient temperature during cooling mode. The device will continue to indicate that the unit is cooling because the compressor will continue to run, however water in the reservoir will begin to return to ambient temperature. The actual water temperature will be displayed but

Distribution pattern

The product was distributed to one customer in MI, and internationally (Taiwan)

Code information

Model Number 222S; Serial #s: 133-2-00697, 133-2-00698; 133-2-00699; 133-2-00700; 133-2-00701; 133-2-00702; and 133-2-00703:.Part # 86165

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1679-2014
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Cincinnati Sub-Zero Products Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1679-2014. Report a data issue: hello@regdatalab.com.