Z-1685-2021 Class II Terminated
Recalled by Siemens Medical Solutions USA, Inc
— Malvern, PA
FDA device recall Z-1685-2021 was initiated by Siemens Medical Solutions USA, Inc on Apr 20, 2021 and is designated Class II. Reason for recall: Siemens identified two issues, In some cases, it is possible to perform an image acquisition although more than 20% of the beam does not hit the detector. This may only occur when an organ program… The recall status is terminated (terminated Nov 8, 2023). Affected quantity: 26.
Recall details
- Product Type
- Devices
- Report Date
- Jun 9, 2021
- Initiation Date
- Apr 20, 2021
- Termination Date
- Nov 8, 2023
- Voluntary/Mandated
- FDA Mandated
- Product Quantity
- 26
Product description
Multitom RAX with software version VF10
Reason for recall
Siemens identified two issues, In some cases, it is possible to perform an image acquisition although more than 20% of the beam does not hit the detector. This may only occur when an organ program for the wall mode without top alignment is selected and the system is moved to the centred position. If afterwards the operator activates the top alignment, increases the collimation, and lifts the tube, the system allows examination even with the tube being misaligned to the detector position. Hence, it is possible to overshoot the detector by more than 20%. From a clinical point of view, it is highly unlikely for the operator to trigger the examination as the misalignment is visible due to the light field being out of range. AND In very rare cases it may occur that the calculated dose value exceeds the limit of 2 Gy. This may potentially occur only when an organ program for free exposure is selected, and the system uses incorrect (too small) source-to-image distance (SID) for calculating applied dose. Hence, the calculated dose is much higher than the actual applied dose causing the buzzer, which normally notifies the operator about the exceeded limit of 2 Gy, to get activated erroneously.
Distribution pattern
US Nationwide distribution
Code information
software version VF10 model# 10860700
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1685-2021
- openFDA file date
- Nov 27, 2026
- Loaded by RegDataLab
- Nov 28, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Siemens Medical Solutions USA, Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Siemens Medical Solutions USA, Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about Siemens Medical Solutions USA, Inc →
Use it in ChatGPT
Get a free API key
How the connector works