Z-1687-2019 Class II Terminated

Recalled by Abbott Vascular — Temecula, CA

Recall Details

Product Type
Devices
Report Date
June 5, 2019
Initiation Date
May 1, 2019
Termination Date
December 11, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
N/A

Product Description

Abbott MitraClip XTR Clip Delivery System, UDI: 08717648226366, Part: CDS0601-XTR, Sterile EO, Rx Only

Reason for Recall

Reports of implantable mitral valve repair system clips unexpectedly opening and becoming nonfunctional, resulting from unintended excessive force applied during the Clip implantation. Excessive force can also result in unexpected movement of clip arms. The inability to close and remove the device has lead to surgery and additional intervention.

Distribution Pattern

U.S.: FL, UT, IN, OK, TX, CO, LA, AZ, MO, AL, GA, CA, AR, KS, IL, HI, IA, KY, MA, DC, NY, NV, NM, NJ, MT, MD, NC, CT, MI, MN, OH, ND, NH, OR, PA, WI, WA, VA, WV, TN, SC, RI, DE, ME, NE, SD, MS, ID, VT. Foreign (OUS): AUSTRALIA, AUSTRIA, BELGIUM, BULGARIA, Canada, CROATIA, CZECH REPUBLIC, DENMARK, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, HUNGARY, INDONESIA, IRELAND, ISRAEL, ITALY, KUWAIT, LEBANON, MACAU, NETHERLANDS, NEW ZEALAND, NORWAY, POLAND, PORTUGAL, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SPAIN, SWEDEN, SWITZERLAND, THAILAND, TURKEY, UNITED KINGDOM

Code Information

All Serial Numbers.