Z-1688-2015 Class II Terminated

Recalled by Terumo Cardiovascular Systems Corporation — Ann Arbor, MI

FDA device recall Z-1688-2015 was initiated by Terumo Cardiovascular Systems Corporation on May 1, 2015 and is designated Class II. Reason for recall: Terumo CVS is implementing field correction activities to address the identified causes of the Terumo System 1 experiencing loss of system power. These activities will be implemented in a phased… The recall status is terminated (terminated Sep 28, 2018). Affected quantity: 1713 units total.

Recall details

Product Type
Devices
Report Date
Jun 3, 2015
Initiation Date
May 1, 2015
Termination Date
Sep 28, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1713 units total

Product description

Base for Terumo Advanced Perfusion System 1, 220/240V The Terumo¿ Advanced Perfusion System 1 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment.

Reason for recall

Terumo CVS is implementing field correction activities to address the identified causes of the Terumo System 1 experiencing loss of system power. These activities will be implemented in a phased approach beginning late 2015: - Implementing design changes and upgrades to improve the reliability of power switching and battery backup. - Adding a switch protector to the On/Off power switch to preven

Distribution pattern

Worldwide Distribution-US (nationwide) including DC and the states of AL , AR , AZ , CA , CO , CT , DE , FL , GA , HI , IA , ID , IL , IN , KY , LA , MA , MD , ME , MI , MN , MO , MS , MT , NC , ND , NE , NJ , NM , NV , NY , OH , OK , OR , PA , SC , TN , TX , UT , VA , VT , WA , WI , and WV, and the countries of Mexico , AUSTRALIA , UNITED ARAB EMIRATES (UAE) , Hong Kong , Indonesia , Singapore , Taiwan , Thailand , Philippines , South Korea , CHILE , Vietnam , Malaysia , BELGIUM , Japan , and CANADA.

Code information

Catalog number: 801764 and All serial number units

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1688-2015
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Terumo Cardiovascular Systems Corporation? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Terumo Cardiovascular Systems Corporation: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Terumo Cardiovascular Systems… →

Source: openFDA record for Z-1688-2015. Report a data issue: hello@regdatalab.com.