Z-1689-2019 Class II Terminated

Recalled by Exactech, Inc. — Gainesville, FL

Recall Details

Product Type
Devices
Report Date
June 5, 2019
Initiation Date
April 26, 2019
Termination Date
April 22, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2 units

Product Description

Opteform Allograft Disc, 45mm x 5mm, 8cc, Catalog Number 600-03-45

Reason for Recall

The Opteform Disc-30mm, Opteform Disc-45mm, and Opteform Disc-90mm were potentially exposed to higher temperatures than those documented in the IFU due to a failure of the cold storage equipment.

Distribution Pattern

OK, VA, FL

Code Information

UDI 10885862093219 Serial Numbers: T32309666, T32309672