Z-1690-2026 Class II Ongoing
FDA device recall Z-1690-2026 was initiated by Medline Industries, LP on Feb 13, 2026 and is designated Class II. Reason for recall: Connectors were not manufactured to required dimensional specifications and may not form a proper seal with ENFit-style devices, which could lead to leakage. Leakage may result in delayed care and… The recall status is ongoing. Affected quantity: 651,789.
Recall details
- Product Type
- Devices
- Report Date
- Apr 8, 2026
- Initiation Date
- Feb 13, 2026
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 651,789
Product description
ENFIT G-TUBE CONNECTOR, MEDLINE Item No. ENFIT1010GC
Reason for recall
Connectors were not manufactured to required dimensional specifications and may not form a proper seal with ENFit-style devices, which could lead to leakage. Leakage may result in delayed care and fluid loss. Additional consequences may also occur if user over-tightens the connector to stop the leakage, including additional clinical intervention, risk of infection, patient discomfort, and prolonged recovery process.
Distribution pattern
Worldwide - US Nationwide and the countries of Canada.
Code information
UDI-DI 10888277314719 (ea) 40888277314710(case) All Lots