Z-1690-2026 Class II Ongoing

Recalled by Medline Industries, LP — Northfield, IL

FDA device recall Z-1690-2026 was initiated by Medline Industries, LP on Feb 13, 2026 and is designated Class II. Reason for recall: Connectors were not manufactured to required dimensional specifications and may not form a proper seal with ENFit-style devices, which could lead to leakage. Leakage may result in delayed care and… The recall status is ongoing. Affected quantity: 651,789.

Recall details

Product Type
Devices
Report Date
Apr 8, 2026
Initiation Date
Feb 13, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
651,789

Product description

ENFIT G-TUBE CONNECTOR, MEDLINE Item No. ENFIT1010GC

Reason for recall

Connectors were not manufactured to required dimensional specifications and may not form a proper seal with ENFit-style devices, which could lead to leakage. Leakage may result in delayed care and fluid loss. Additional consequences may also occur if user over-tightens the connector to stop the leakage, including additional clinical intervention, risk of infection, patient discomfort, and prolonged recovery process.

Distribution pattern

Worldwide - US Nationwide and the countries of Canada.

Code information

UDI-DI 10888277314719 (ea) 40888277314710(case) All Lots

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1690-2026
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1690-2026. Report a data issue: hello@regdatalab.com.