Z-1693-2015 Class II Terminated

Recalled by Neusoft Medical System Co — Houston, TX

Recall Details

Product Type
Devices
Report Date
June 24, 2015
Initiation Date
March 8, 2014
Termination Date
December 5, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5 installed in US

Product Description

Neusoft NeuViz 64 Multi-slice CT Scanner System

Reason for Recall

The A-plane compensator is made of Teflon which will become fragile due to X-ray exposure passing through it. The thinnest part of the compensator is subject to radiation degradation and as a result has been found to fail prematurely. Failure is indicated by pin or ring artifact in the image

Distribution Pattern

Worldwide Distribution - US Distribution including Puerto Rico and to the states of : IL, MA, TX and SC., and to the countries of : China, Malaysia, Lebanon, Ecuador, Egypt, United Arab Emirates,Kazashstan, Tunisia, Angola, Brazil, Uzbekistan, Lebanon, Cote D' ivoire, South Africa, Montenegro and Tajikistan.

Code Information

NeuViz 16/64 Series