Z-1693-2015 Class II Terminated

Recalled by Neusoft Medical System Co — Houston, TX

FDA device recall Z-1693-2015 was initiated by Neusoft Medical System Co on Mar 8, 2014 and is designated Class II. Reason for recall: The A-plane compensator is made of Teflon which will become fragile due to X-ray exposure passing through it. The thinnest part of the compensator is subject to radiation degradation and as a result… The recall status is terminated (terminated Dec 5, 2016). Affected quantity: 5 installed in US.

Recall details

Product Type
Devices
Report Date
Jun 24, 2015
Initiation Date
Mar 8, 2014
Termination Date
Dec 5, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5 installed in US

Product description

Neusoft NeuViz 64 Multi-slice CT Scanner System

Reason for recall

The A-plane compensator is made of Teflon which will become fragile due to X-ray exposure passing through it. The thinnest part of the compensator is subject to radiation degradation and as a result has been found to fail prematurely. Failure is indicated by pin or ring artifact in the image

Distribution pattern

Worldwide Distribution - US Distribution including Puerto Rico and to the states of : IL, MA, TX and SC., and to the countries of : China, Malaysia, Lebanon, Ecuador, Egypt, United Arab Emirates,Kazashstan, Tunisia, Angola, Brazil, Uzbekistan, Lebanon, Cote D' ivoire, South Africa, Montenegro and Tajikistan.

Code information

NeuViz 16/64 Series

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1693-2015
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Neusoft Medical System Co? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Neusoft Medical System Co: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Neusoft Medical System Co →

Source: openFDA record for Z-1693-2015. Report a data issue: hello@regdatalab.com.