Z-1695-2019 Class II Ongoing

Recalled by Philips North America, LLC — Andover, MA

FDA device recall Z-1695-2019 was initiated by Philips North America, LLC on Mar 28, 2019 and is designated Class II. Reason for recall: The MX40 may experience increased power consumption and may have a lack of visual and/or audible notifications while in a Low Battery State . The recall status is ongoing.

Recall details

Product Type
Devices
Report Date
Jun 12, 2019
Initiation Date
Mar 28, 2019
Voluntary/Mandated
Voluntary: Firm initiated

Product description

IntelliVue MX40 Patient Monitor, Product number 857146, Service Numbers All: 453564711531 and 453564711541 - Product Usage: Used primarily as a traditional telemetry medical device. It connects to the IntelliVue Information Center by way of a wireless network. When the MX40 is connected to the IntelliVue Information Center the IntelliVue Information Center provides the primary patient monitoring and alarming function. The MX40 does not automatically provide local monitoring or alarming when connected to the Information Center.

Reason for recall

The MX40 may experience increased power consumption and may have a lack of visual and/or audible notifications while in a Low Battery State .

Distribution pattern

Worldwide distribution - US Nationwide and to 71 countries.

Code information

All revision B.X versions are affected.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1695-2019
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Philips North America, LLC? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Philips North America, LLC: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Philips North America, LLC →

Source: openFDA record for Z-1695-2019. Report a data issue: hello@regdatalab.com.