Z-1698-2019 Class II Terminated

Recalled by Microline Surgical, Inc. — Beverly, MA

FDA device recall Z-1698-2019 was initiated by Microline Surgical, Inc. on Mar 22, 2019 and is designated Class II. Reason for recall: The shears exhibited a high rate of complete or intermittent failure to actuate. As a result, they do not perform their intended function in surgical procedures. The recall status is terminated (terminated Jul 30, 2020). Affected quantity: 115 units.

Recall details

Product Type
Devices
Report Date
Jun 12, 2019
Initiation Date
Mar 22, 2019
Termination Date
Jul 30, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
115 units

Product description

TLS2 Thermal Ligating Shears, Catalog Number 132-131D

Reason for recall

The shears exhibited a high rate of complete or intermittent failure to actuate. As a result, they do not perform their intended function in surgical procedures.

Distribution pattern

The products were distributed to the following US states: ND and NY. The products were distributed to the following foreign countries: Germany, Netherlands.

Code information

UDI: 811099011439; Lot Numbers: 807118, 807119, 807120, 811076

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1698-2019
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Microline Surgical, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1698-2019. Report a data issue: hello@regdatalab.com.